Alert Codes
Understand the alert codes on the Vagus.net tVNS Stimulator.
A861
Impossibly High Resistance
Error
Stimulation Cannot Start
Stimulation Cannot Start
- Clip Not Connected
- Impedance Impossibly High
- Check setup
On startup the device tries to pass a very small current to check if an ear is present. If it is not possible then A861 is displayed and the device automatically turns off
BAD SETUP
A862
Very High Resistance
Limited
Mild Stimulation Only
Mild Stimulation Only
- Inspect Setup for source of resistance
- Clean Surfaces
- Apply More Conductive Paste
A862 occurs when there is high resistance between the electrodes.
Easy Fix
A863
Medium Resistance
Almost
Cannot Reach Maximum
Cannot Reach Maximum
- Good Stimulation
- Not Quite Maximum
- Check Setup
A863 indicates there is some resistance preventing maximum stimulation.
Don't Worry
A864
Cable Fault Detected
FAULT
Short Circuit Detected
Short Circuit Detected
- Contact support
- Please send photo of setup
The device has detected a problem on the connected jack. Only connect official Vagus.net cables and clip electrodes.
A865
Empty Battery
EMPTY
Please Recharge Battery
Please Recharge Battery
- Battery empty
- Connect Charging Cable
- To a 5v USB Power Source
- And Recharge Device
- Until Battery Charged
Please recharge the device using a vagus.net USB charging cable. Replacement cables can be ordered here - tvns charge cable
RECHARGE